7
Stand-by Indicator
The
Stand-By
indicator will be ON when the Unit is preparing to produce sanitizer. This will be either during the system’s
initial
Start-Up
or when the ‘Cell’ is in an OFF cycle during the filtration cycle (refer
System Control
). Stand-By indicator
can also be ON after System Protection Cut-Out has activated. (ie: when unit has shut down).
Flow Indicator
If there is a problem with water flow or gas is detected in the ‘Cell Housing’ the
Flow
indicator will be ON. When this
occurs the pump or pipes should be inspected for damage and the
Gas Sensor
on the ‘Cell’ checked for correct
connection (and scale build-up on bolt head of Gas Sensor).
System Control (Output Control)
The
System Control
allows you to control the amount of sanitizer to be produced during any filtration cycle (filter
operating time).
The setting on the
System Control
determines the amount of time for which the ECO-matic will operate during the filtration
cycle. The
System Control
dial is graduated approximately as follows: 0% (MIN), 20%, 40%, 60%, 80%, (MAX) 100%.
The
System Control
will not vary the rate at which the ‘Cell’ will produce sanitizer (as indicated on the DISPLAY), just the
‘time’ for which the ECO-matic will produce sanitizer. For example:
Daily Filter
System Control
Hours of
Operating Time
Setting
Sanitizer Production
8 hours
60%
4.8 hours (60% of 8 hours)
12 hours
20%
2.4 hours (20% of 12 hours)
8 hours
100%
8 hours (100% of 8 hours)
6 hours
50%
(halfway) 3 hours (50% of 6 hours)
When the
System Control
is set to MIN, the ‘Cell’ will be OFF for the duration
of the filtration cycle.
When the
System Control
is set to MAX, the ‘Cell’ will be ON for the full duration
of the filtration cycle.
The
System Control
is graduated in steps of 20% from MIN (OFF) to MAX (ON).
Note: If the ECO-matic is controlled by ORP chemical automation, turn the control knob to MAX.
During any filtration cycle, when the
System Control
is set between MIN and MAX, the ‘Cell’ will be turned ON and OFF
a number of times each hour. Using the previous example (of 60%), the ‘Cell’ will operate for about 36 minutes each hour.
This 36 minute operating time will be made up of a number of smaller operating periods. As an example, the ‘Cell’ may
turn ON 12 times (for a period of 3 minutes each time) to make up the 36 minutes. This enables the electronic circuitry to
re-adjust to any changes in the pool water condition. For example, dilution from winter rains, the addition of salt, etc.
If the ‘Cell’ is OFF and you wish to check its operation, simply turn the
System Control
to MAX and the ’Cell’ will turn ON.
Once checked, adjust the
System Control
back to the desired position and after a few minutes the ‘Cell’ will turn OFF
again. To turn the ‘Cell’ OFF, simply turn the
System Control
to MIN. This will be convenient when backwashing filter.
Summary of Contents for ESC 16
Page 3: ...Installation and Operating Instructions ESC 16 ESC 24 ESC 36 ESC 48 FOR MODELS TM...
Page 24: ...22 Notes...
Page 25: ......
Page 26: ...P N 401633 2 TM...
Page 27: ...ECO matic ESC Chlorinator SERVICING INSTRUCTIONS ESC SERVICE GUIDE...
Page 52: ...26 Troubleshooting Guides...
Page 53: ...27...
Page 54: ...28...
Page 55: ...29...
Page 56: ...30...
Page 64: ...38 NOTES...
Page 65: ...39...
Page 68: ...ESC SERIES SALT CHLORINATOR...
Page 69: ...EcoSalt BMSC Chlorinator SERVICING INSTRUCTIONS ECOSALT SERVICE GUIDE...
Page 89: ...21 Trouble Shooting Guides...
Page 90: ...22...
Page 91: ...23...
Page 92: ...24...
Page 97: ...EcoSpa ORP Controlled Sanitising System SERVICING INSTRUCTIONS ECOSPA SERVICE GUIDE...
Page 109: ...13 Step 4 Place cable ties into position holding the fuse wires in place...
Page 112: ...16 Trouble Shooting Guides...
Page 113: ...17...
Page 114: ...18...
Page 115: ...19...
Page 116: ...20...
Page 117: ...21 This page intentionally blank...
Page 118: ...22 NOTES...
Page 119: ...23 NOTES...