5-16
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Message: V Tach
Parameter value: VT
n
or more consecutive PVCs have
been detected with a beat-to-beat
rate greater than or equal to the VT
rate set in the Alarm Tables. (
n
=
VT count parameter)
• Observe the patient and treat if
clinically indicated.
Message: Bradycardia
Parameter value:
BRDY
8 or more consecutive QRS
complexes have been detected
with a beat-to-beat rate less than
80% of the patient’s normal rate.
• Observe the patient and treat if
clinically indicated.
Message: RUN
Parameter value: RUN
3 to N-1* consecutive PVCs with a
beat-to-beat rate >= the VT rate
• Observe the patient and treat if
clinically indicated.
Message: AIVR
Parameter value:
AIVR
3 or more PVCs with a rate < the
VT rate
• Observe the patient and treat if
clinically indicated.
Message: SVT
Parameter value: SVT
N* or more consecutive normal
beats with a beat-to-beat rate >=
the SVT rate
• Observe the patient and treat if
clinically indicated.
Message: CPT
Parameter value: CPT
A sequence of beats with the
pattern: normal, PVC, PVC, normal
• Observe the patient and treat if
clinically indicated.
Message: BGM
Parameter value: BGM
A sequence of beats with the
pattern: normal, PVC, normal,
PVC, normal
• Observe the patient and treat if
clinically indicated.
Message: TACH
Parameter value:
TACH
N* or more consecutive normal
beats with a beat-to-beat rate >=
the TACH rate
• Observe the patient and treat if
clinically indicated.
Message: PAUS
Parameter value:
PAUS
A sequence of two normal beats
with an N-N* interval > the Pause
rate times the average N-N interval
(±100ms)
• Observe the patient and treat if
clinically indicated.
Message: ST <lead>
Off
ST<lead1> and
ST<lead2> parameter
value: ***
The monitor has detected a lead-off
condition for the currently
processed lead. The cause could
be one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the electrode
Wrong cable type selection.
The M
ULTIMED
/N
EOMED
pod may
be defective.
• Check the patient cable and lead
wires carefully.
• Check the M
ULTIMED
pod.
Replace pod if defective.
• Replace any cable or lead wire
that is suspect.
• Reapply gel on the electrode or
change the electrode (if
disposable).
• Select another ECG lead for
processing if the electrode or lead
cannot be replaced.
Heart Rate, Arrhythmia, ST Segment Analysis
(continued)
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...