5-6
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
For the AutoSet function, the monitor calculates alarm limits as a percentage of the
current parameter values as follows:
Turning Parameter Alarms On/Off
The monitor annunciates alarms only for parameters whose alarm is turned on. You
can turn alarms on or off for each parameter individually.
When you turn a parameter alarm off, the monitor displays a crossed bell icon in the
corresponding parameter box.
STEPS: Turning Parameter Alarms On or Off
1. Press the
Alarm Limits
fixed key.
2. Scroll to the
Alarm
column of the desired parameter and click the knob.
3. Dial in the desired setting and click the knob.
Parameter
Upper Limit
Lower Limit
SpO2
fixed at 100%*
-5%
HR, PLS, Rsp, NBP, GP1, GP2
+20%
-15%
Tempe7%
-7%
ST
+2.0
-2.0
*For neonatal monitoring mode, the Upper Limit Adjustment is fixed at 95%.
Higher adjustments must be made manually.
WA R N I N G : I f t h e m o n i t o r d i s p l a y s t h e b a n n e r
H R , A S Y, V F
O ff
, t h e r e i s n o a l a r m a n n u n c i a t i o n f o r a s y s t o l e a n d
v e n t r i c u l a r f i b r i l l a t i o n .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...