R
ELEARNING
A
P
ATIENT
’
S
ECG
VF6
Infinity Vista
9-7
Arrhythmia Alarm Recordings
STEPS: Turning Alarm Recordings On/Off
1. Call up the
Arrhythmia Setup
table (see above).
2. Click on the recording setting in the
Record
column of the desired arrhythmia
parameter.
3. Dial in the desired setting (
Record
,
Store
,
Str/Rec
, or
OFF
).
Relearning a Patient’s ECG
It is advisable to initiate a relearning under one or more of the following conditions:
z
Eight hours have passed since the last learning of a reference complex.
z
You repositioned the electrodes.
z
You observe clinically questionable arrhythmia calls.
z
There have been significant changes in the patient’s ECG.
z
The message
ARR
Relearn
appears in the message area.
The monitor automatically relearns in the following cases:
z
When you turn arrhythmia monitoring on or change between basic and full
arrhythmia monitoring.
z
When you change the displayed ECG leads while arrhythmia monitoring is
on.
NOTE:
z
You cannot turn alarm recordings for asystole (ASY) and ventricular fibrillation
(VF) off. For these events, only the settings
Record
and
Str/Rec
are available.
z
See the chapter
Recordings
for a description of alarm and stored recordings and
how to view, save, print, or delete stored recordings via the Event Recall screen.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...