4-4
Infinity Vista
VF6
4 A
DMISSION
/D
ISCHARGE
/T
RANSFER
Patient Category
STEPS: Selecting the Patient Category
1. Call up the Patient Admit menu (Menu > Admit/Discharge > Patient Admit,
see above).
2. Click on
Patient Category
.
3. Dial in the desired patient category (Adult, Pediatric, Neonate) and click the
knob.
When you change the patient category, the monitor:
z
Clears all current alarms, including their messages.
z
Stops any NBP measurement in progress.
z
Deletes stored trend data, events, and recordings.
z
Returns alarm limits to their category-specific default settings (see appendix
Default Settings and Biomedical Support
).
z
Returns waveform display scales (sizes) to their category-specific default
settings (see appendix
Default Settings and Biomedical Support
).
z
Sets the display range for trend values according to the selected patient
category (see
Trends
chapter).
NOTE:
The currently selected patient category is indicated between the first and
second waveform channels next to the parameter boxes.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...