I
MPORTANT
G
ENERAL
S
AFETY
C
ONSIDERATIONS
VF6
Infinity Vista
1-5
Safety, Inspection, and Maintenance
In the interest of safety, regular equipment inspection and maintenance is required.
Once a year, check all cables, devices, and accessories for damage, ground resistance,
chassis and patient leakage currents, and all alarm functions. Also, ensure that all
safety labels are legible. Maintain a record of these safety checks. For additional
information, refer to the Service manual.
Do not use cables that appear cracked, worn or damaged. Use of such cables may
contribute to poor monitoring performance or cause the display of erroneous values.
Leakage current increase when connecting multiple medical devices to a patient.
Ensure the electrical shock classification for each device is suitable for the intended
application.
Dräger recommends that safety and functional checks be performed on the monitor
at
least
once each year. The temperature and non-invasive blood pressure circuits of the
monitor should be calibrated at least every two years. These checks should be
performed by authorized personnel, as described in the appropriate Service manual.
When main or battery power is not available, the monitor stores patient data and
settings in an internal battery backed-up SRAM. This internal battery lasts
approximately 10 years if the monitor is operated from main power or from the lead
acid or lithium ion battery.
WA R N I N G : B e c a u s e o f t h e d a n g e r o f e l e c t r i c s h o c k , n e v e r
r e m o v e t h e c o v e r o f a n y d e v i c e w h i l e i n o p e r a t i o n o r
c o n n e c t e d t o a p o w e r o u t l e t v i a t h e A C a d a p t e r.
CAUTION:
To preserve the life of the internal battery, always leave the monitor
connected to main power (using the AC adapter) when not in use. If the monitor is
stored unconnected from line or lead acid/lithium ion battery power, the capacity of the
internal battery is drained in approximately three years.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...