NBP M
EASUREMENTS
VF6
Infinity Vista
13-7
STEPS: Taking NBP Interval Measurements
1. Press and hold the
NBP Start/Stop
key,
or
1. Click on the
NBP
parameter box.
2. Click on
Interval Mode
.
3. Select the desired interval and click the knob.
To end interval measurements, press and hold the
NBP Start/Stop
key or turn the
Interval Mode
OFF
in the NBP menu.
NOTE:
z
When you press and hold the NBP Start/Stop key to turn the Interval Mode on,
the interval time is the one that was previously selected.
z
If the Interval Mode is active before you power cycle the monitor, the Interval
Mode setting is retained through a power cycle. If the Interval Mode is turned
off before you power cycle the monitor, you have to reselect an interval time in
the NBP menu, before interval measurements can be started with the NBP Start/
Stop key.
z
After exiting Standby, you have to restart NBP measurement cycles by pressing
the NBP Start/Stop key.
z
During an active single NBP measurement, you cannot turn the Interval Mode
on or off.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...