
A
LARM
S
ETTINGS
VF6
Infinity Vista
5-5
Setting Alarm Limits
You can set alarm limits for each parameter individually, or you can use the AutoSet
function to set alarm limits for all parameters at once based on the current parameter
values.
STEPS: Setting alarm limits individually
1. Press the
Alarm Limits
fixed key.
2. Scroll to the
Upper
column for the desired
parameter and click the knob.
3. Dial in the desired upper limit and click the knob.
4. Scroll to the
Lower
column and click the knob.
5. Dial in the desired lower limit and click the knob.
STEPS: Using the AutoSet function
1. Press the
Alarm Limits
fixed key.
2. Scroll to
AutoSet
. The column AutoSet shows
the upper and lower limits that are automatically
calculated for each parameter based on current
parameter values.
3. To accept these limits for all parameters, click on
AutoSet
. The automatically calculated limits are
written into the Upper and Lower columns.
NOTE:
z
The AutoSet function sets alarm limits for all parameters at once, even for those
not displayed on the current page of the Alarm Limits table. Exception: There is
no AutoSet function for PVC.
z
If a parameter value is not currently available (e.g. during a lead-off condition),
the monitor does not adjust the alarm limit for that parameter when you click on
AutoSet.
z
After using the AutoSet function, you can change individual alarm limits at any
time.
z
Due to slight rounding differences in values, the alarm limits display for kPa
values at the central station does not match the alarm limits display for kPa
values at the Infinity Vista monitor.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...