Arrhythmia Monitoring Modes
Label
Event
Description
OFF: The monitor detects these events when Arrhythmia = OFF (via ECG)
ASY
Asystole
4 seconds pass without the detection of a valid QRS
complex
VF
Ventricular
Fibrillation
The monitor identifies a sinusoidal waveform with
fibrillation characteristics
Basic: The monitor detects these additional events when Arrhythmia = Basic
VT
Ventricular
Tachycardia
N* or more consecutive PVCs with a beat-to-beat rate >=
the VT rate
BRDY**
Sinus
Bradycardia
8 or more consecutive normal beats with an average rate
<= the BRDY rate
Full: The monitor detects these additional events when Arrhythmia = Full
RUN
Ventricular
Run
3 to N-1* consecutive PVCs with a beat-to-beat rate >= the
VT rate
AIVR
Accelerated
Idioventricular Rhythm
3 or more PVCs with a rate < the VT rate
SVT
Supraventricular
Tachycardia
N* or more consecutive normal beats with a beat-to-beat
rate >= the SVT rate
CPT
Ventricular
Couplet
A sequence of beats with the pattern: normal, PVC, PVC,
normal
BGM
Ventricular
Bigeminy
A sequence of beats with the pattern: normal, PVC,
normal, PVC, normal
TACH
Sinus
Tachycardia
N* or more consecutive normal beats with a beat-to-beat
rate >= the TACH rate
PAUS
Pause
A sequence of two normal beats with an N-N* interval >
the Pause rate times the average N-N interval (±100ms)
ARTF
Artifact
A sinusoidal signal with a frequency of 10 to 20 Hz and an
amplitude > 0.4 mV p-p, or with a frequency of 0.25 to
10 Hz and an amplitude > 10 mV p-p
*
N is the event count set on the Arrhythmia Setup Table.
**
In neonatal mode, you can set alarm limits for BRDY in the Alarm Limits table and the
monitor alarms upon a limit violation for this event.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...