11-6
Infinity Vista
VF6
11 R
ESPIRATION
M
ONITORING
Resp Display Amplitude
The
Size
option lets you increase or decrease the amplitude of the respiration
waveform. In manual breath detection sensitivity mode (see the section
Selecting the
Rsp Mode
, below), the detection sensitivity is based on the selected waveform
amplitude. Available amplitude settings are:
z
10% to 100%, in increments of 10% (default 50%)
STEPS: Selecting the Rsp Display Amplitude
1. Click on the
Rsp waveform
channel.
2. Click on
Size
.
3. Select the desired display amplitude and click the knob.
WA R N I N G : I n M a n u a l M o d e , i f y o u s e t t h e s i z e o f t h e
r e s p i r a t i o n w a v e f o r m t o o l o w, s h a l l o w b r e a t h s m a y n o t b e
c o u n t e d . I f s e t t o o h i g h , c a r d i a c a r t i f a c t m i g h t b e c o u n t e d a s
b r e a t h s . T h e r e f o r e , a l w a y s t u r n r e s p i r a t i o n m a r k e r s o n i n
m a n u a l m o d e t o h e l p y o u s e l e c t a n a p p r o p r i a t e w a v e f o r m
a m p l i t u d e / d e t e c t i o n s e n s i t i v i t y ( s e e t h e s e c t i o n
D i s p l a y i n g
R s p M a r k e r s
) .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...