
ECG
AND
HR S
AFETY
C
ONSIDERATIONS
VF6
Infinity Vista
8-15
ECG and HR Safety Considerations
HR Alarm Settings
Set HR alarm limits on the Alarm Limits table (see the chapter
Alarms and Messages
).
If you turn the HR alarm off, the monitor displays a crossed bell icon in the HR
parameter box and the banner
HR Limits Off
above the first waveform channel. If
arrhythmia monitoring and alarms for asystole and ventricular fibrillation are also
disabled, the monitor displays the banner
HR, ASY, VF, Off
.
Neonatal ECG Monitoring
To obtain the most accurate heart rate count and waveform display, select the ECG
lead(s) with the smallest T wave and the largest monophasic QRS.
ECG 50/60 Hz Notch Filter Setting
When the line frequency is set incorrectly, the resulting 50 or 60 Hz noise on the ECG
signal can make the ECG waveform unreadable along with masking QRS complexes.
As a result, detection of the QRS complexes becomes unreliable. When these cannot
be sensed reliably, incorrect HR calculations may result, leading to both false positive
and false negative HR alarms.
Muscle Stimulators
NOTE:
z
If arrhythmia detection is on, alarms for asystole (ASY) and ventricular
fibrillation (VF) are automatically on and cannot be turned off.
z
By default, alarms for asystole and ventricular fibrillation are always on;
however, these alarms can be disabled in the password-protected Biomed menu
(see the appendix
Default Settings and Biomedical Support
).
WA R N I N G : I f t h e m o n i t o r d i s p l a y s t h e b a n n e r
H R , A S Y, V F
O ff
, t h e r e i s n o a l a r m a n n u n c i a t i o n f o r a s y s t o l e a n d
v e n t r i c u l a r f i b r i l l a t i o n .
WA R N I N G : N e o n a ta l Q R S c o m p l e x e s t h a t a r e b i p h a s i c a n d
l e s s t h a n 4 0 m s w i d e , o r m o n o p h a s i c a n d l e s s t h a n 2 0 m s
w i d e , a r e d i s p l a y e d s m a l l e r t h a n t h e i r a c t u a l a m p l i t u d e .
WA R N I N G : T h e u s e o f e l e c t r i c a l m u s c l e s t i m u l a t o r s m a y
i n t e r f e r e w i t h E C G a n d r e s p i r a t i on m o n i t o r i n g . I f t h i s o c c u r s ,
d i s c o n t i n u e t h e i r u s e .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...