Electrosurgery (ESU)
The monitor’s ECG function is protected against high-frequency interference from
defibrillators and electrosurgical units. The ESU B
LOCK
(optional) enhances the
performance of the monitor during electrosurgery and allows you to use standard ECG
lead sets. It reduces noise on ECG tracings and helps protect the patient from burns.
STEPS: Using the ESU Block
1. Turn off pacemaker detection.
2. Plug the ESU block into the M
ULTI
M
ED
5 pod as shown above. If you are
using a M
ULTI
M
ED
6 pod, ignore the extra V lead connection.
3. Plug a standard white ECG lead set into the ESU block as shown. Do not use
shielded blue lead wires with the ESU block.
4. Remove the ESU block to continuously monitor the patient.
WA R N I N G :
z
D r ä g e r r e c o m m e n d s u s i n g t h e E S U b l o c k d u r i n g
e l e c t r o s u r g e r y. I f y o u d o n o t h a v e a n E S U b l o c k , u s e
o n l y D r ä g e r b l u e E C G l e a d w i r e s . T h e y h e l p p r o t e c t t h e
pa t i e n t f r o m b u r n s c a u s e d b y E S U - i n d u c e d c u r r e n t
f l o w i n g t h r o u g h t h e l e a d w i r e s .
z
I m p e d a n c e r e s p i r a t i o n m o n i t o r i n g a n d pa c e m a k e r s p i k e
d e t e c t i o n a r e i n o p e r a t i v e w h e n u s i n g t h e E S U b l o c k .
z
T h e N
E O
M
E D
p o d i s n o t i n t e n d e d f o r u s e d u r i n g
e l e c t r o s u r g e r y. To p r o t e c t pa t i e n ts f r o m b u r n s , d o n o t
u s e t h e N
E O
M
E D
p o d i n a n E S U e n v i r o n m e n t .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...