C-9
Checking the NBP Calibration
The
NBP
parameter box gives you access to the NBP Calibration mode. These checks
must be performed in Adult Mode only.
STEPS: NBP Calibration
1. Click on the
NBP
parameter box.
2. Click on
Calibration Mode
.
3. Select
ON
and click the knob.
4. Replace the cuff with a mercury manometer.
5. Pump the manometer until the mercury column reads 260 mmHg. Verify the
displayed value on the monitor.
6. Release the pressure and read the pressure value at three different points on
the mercury column: the displayed values on the monitor should be within ±3
mmHg of the manometer readings.
7. Close the release valve on the bulb and pump the manometer until the
overpressure release valve opens. This should happen at 300 mmHg ±30
mmHg and the mercury column should fall rapidly to zero. If the release valve
does not open or if it opens at a value lower than 300 mmHg ±30 mmHg, the
overpressure system may be faulty. Remove the monitor from service and
contact DrägerService.
NOTE:
If the monitor is left in the calibration mode for more than 128 seconds after a
first measurement has been started, an
NBP Fault
condition is reported
.
This message
does not indicate a hardware failure but the activation of the safety timer of the monitor.
To clear the message, turn the monitor off, then on again.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...