M
ESSAGES
VF6
Infinity Vista
5-21
Message:
SpO
2
Searching
Parameter value:
<blank>
The monitor is analyzing the signal.
This occurs during the first 5 to 10 s
of operation or as a result of motion
artifact.
• Wait until the message
disappears from the screen.
Message: SpO
2
Transparent
Message: PLS
Transparent
Parameter value: ***
The sensor may be disconnected
from the patient.
Sensor may be defective.
• Check sensor and connections.
• If problem persists, change
sensor.
Message: SpO
2
Unplugged
Message: PLS
Unplugged
Parameter value: ***
The sensor’s extension cable
disconnected from the pod.
The sensor is disconnected from
the extension cable.
The M
ULTIMED
/N
EOMED
cable is
defective.
The sensor is defective.
The sensor is not recommended
for use with your monitor.
• Verify sensor to extension cable
connection.
• Verify extension cable to pod
connection.
• Verify pod connection to monitor.
• Verify correct sensor.
Message: Replace
SpO
2
Sensor
The SpO
2
sensor is not functioning
properly.
• Replace the sensor.
Message: SpO
2
<LL
<value>
Message: SpO
2
>UL
<value>
Parameter value:
<value>
Message: PLS<LL
<value>
Message: PLS>UL
<value>
Parameter value:
<value>
The patient’s oximetry level or
pulse rate falls outside the current
alarm limits, or the alarm limits are
inappropriate for this patient.
The SpO
2
level has changed since
you last selected the alarm limits.
Elevated carboxyhemoglobin or
methemoglobin levels. Significant
levels of intravascular dyes.
Placement of sensor on an
extremity that has a blood pressure
cuff, arterial catheter or
intravascular line.
Faulty equipment.
• Observe the patient.
• Apply treatment if clinically
indicated.
• Change the alarm limits.
• If you suspect a sensor failure,
check and replace the sensor, if
necessary.
Message:
SpO
2
: non-Masimo
sensor
SpO
2
: non-Nellcor
sensor
Parameter Value: ***
Sensor not compatible with monitor
configuration.
• Use compatible sensor (Masimo/
Nellcor).
• Contact your hospital’s
Biomedical department to change
the monitor configuration.
Message:
SpO
2
: Duplicate
Device Connected
A MultiMed with an SpO
2
sensor
intermediate cable
and
an SpO
2
pod are connected to the monitor at
the same time.
• Remove the sensor intermediate
cable from the MultiMed; or
• Remove the SpO
2
pod.
Pulse Oximetry (SpO2)
(continued)
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...