9-6
Infinity Vista
VF6
9 A
RRHYTHMIA
Rate and Count
STEPS: Selecting Rate and Count Limits
1. Call up the
Arrhythmia Setup
table (see above).
2. Click on the limit in the
Rate
or
Count
column for the desired arrhythmia
parameter.
3. Dial in the desired limit and click the knob.
Arrhythmia Alarms
STEPS: Turning Arrhythmia Alarms On/Off
1. Call up the
Arrhythmia Setup
table (see above).
2. Click on the alarm setting in the
Alarm
column of the desired arrhythmia
parameter.
3. Dial in the desired setting (
ON/OFF
) and click the knob.
NOTE:
You can change limits only for VT, BRDY (rate only), SVT, TACH, and
PAUS (rate only). The displayed limits for RUN and AIVR are informational.
NOTE:
You cannot turn the alarms for asystole (ASY) and ventricular fibrillation
(VF) off in the Arrhythmia Setup table. Disabling alarms for these life-threatening
events is a password-protected function (see
Default Settings and Biomedical Support
).
WA R N I N G : C e r ta i n v e n t r i c u l a r ta c h y c a r d i a ( V T ) h a v e
s i n u s o i d a l w a v e f o r m s c l o s e l y r e s e m b l i n g t h o s e o f v e n t r i c u l a r
f i b r i l l a t i o n ( V F ) . B e c a u s e o f t h e s i m i l a r i t y o f t h e s e
w a v e f o r m s , t h e m o n i t o r m a y c l a s s i f y s u c h t y p e s o f
v e n t r i c u l a r ta c h y c a r d i a a s v e n t r i c u l a r f i b r i l l a t i o n , t h e m o r e
s e r i o u s o f t h e t w o c o n d i t i o n s .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...