5-14
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Heart Rate, Arrhythmia, ST Segment Analysis
Message
Possible Cause
Suggested Action
Message: Arrhythmia
Relearning
Parameter value: LRN
The monitor is learning the
patient’s normal QRS complex to
use as reference.
• Wait until the message
disappears from the screen.
Message: Asystole
Parameter value: ASY
No QRS detection for the last 4
seconds.
Heart rate is below 15 beats/
minute.
• Observe the patient and treat if
clinically indicated.
• Select a lead with at least an
amplitude of 0.5 mV.
• Select another lead to meet the
above criterion.
• Reposition or change the
electrodes if the amplitude is still
low.
• Use good skin preparation.
Message: RL Lead Off
*
HR parameter value:
<value>
ST<lead1>/ST<lead2>
parameter value:
<blank>
Message: V Lead Off *
Message: V+ Lead Off
Message: RA Lead Off
Message: LL Lead Off
Message: LA Lead Off
Message: ECG Leads
Off
HR parameter value:
***
ST<lead1>/ST<lead2>
parameter value:
<blank>
The monitor has detected a lead-off
condition for the currently
processed lead. The cause could
be one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the electrode
Wrong cable type selection.
The M
ULTIMED
/N
EOMED
pod may
be defective.
• Check the patient cable and lead
wires carefully.
• Check the M
ULTIMED
/N
EOMED
pod. Replace pod if defective.
• Replace any cable or lead wire
that is suspect.
• Reapply gel on the electrode or
change the electrode (if
disposable).
• Select another ECG lead for
processing if the electrode or lead
cannot be replaced.
* For all the electrode disconnected messages (V, V+ RA, LL, LA, RL) the parameter box shows
*** if this is the lead currently being monitored. If another valid lead is being monitored (i.e. ECG
1&2) then the parameter value for HR, ST<lead1> and ST<lead2> is <value>
Message: ARR
Cannot Learn Lead
<lead>
Parameter value:
<value>
Lead not connected.
• Check lead connections
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...