O
VERVIEW
VF6
Infinity Vista
D-9
Resolution
1 breath per minute
Measurement Range
0 to 155 breaths per minute
Measurement Accuracy
± 1 breath per minute, or ± 2%, whichever is greater
Response Time
After a step change in respiration rate, it takes 36 seconds to
stabilize to a new respiration rate of 10 breaths per minute
and 18 seconds to stabilize to a new respiration rate of 20
breaths per minute.
Report Interval
2 seconds
Total resistance
0 to 3600
:
Input Dynamic Range
AC: ± 17.3
:
± 10%
DC: 0
:
to 3.6 k
:
(internal check for high impedance)
Bandwidth
0.25 Hz to 3.5 Hz
Impedance Measuring Current
65
P
A RMS ±10% nominal at 48kHz ±1%, rectangular wave
Pulse Oximetry (SpO
2
)
Parameter Display
Saturation (%SpO
2
), pulse rate
Measuring Method
Absorption-spectrophotometry
Measuring Range
SpO
2
:
1 - 100%
Pulse rate:
30 - 250 1/min
Calibration Range
70-100%
Display Range
0-100%
Display Update Period
2 seconds
Maximum Hold from Previous
Update
30 seconds (in the event of artifact or other error)
Respiration
(continued)
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...