14-4
Infinity Vista
VF6
14 I
NVASIVE
B
LOOD
P
RESSURE
Display Channel and Waveform Amplitude
The monitor displays pressure waveforms in the lower display channels. Depending
on the pressure label, you can select waveform amplitudes within the following
ranges:
STEPS: Selecting the IBP Display Channel and Waveform
Amplitude
1. Click on the desired waveform channel.
2. Click on
Waveform
.
3. Select the desired pressure parameter and click the knob.
4. Click on
Size
.
5. Select the desired display amplitude and click the knob.
Pressure
Label
Possible Size Settings
Adult/Pediatric
Default Setting
Neonatal Default
Setting
ART, GP1, GP2 50, 75, 100,...300 mmHg
(8, 12, 16,...40 kPa)
200 mmHg
(24 kPa)
100 mmHg
(16 kPa)
PA
20, 40, 50, 75, 100, 150 mmHg
(4, 6, 8, 12, 16, 20 kPa)
50 mmHg
(8 kPa)
50 mmHg
(8 kPa)
CVP
-5, 10, 20, 30, 40, 50 mmHg
(-1.0, 2, 4, 5, 6, 8 kPa)
20 mmHg
(4 kPa)
20 mmHg
(4 kPa)
ICP
0, 5, 10, 15, 20, 50, 100 mmHg
(1, 2, 3, 4, 8, 16 kPa)
20 mmHg
(4 kPa)
10 mmHg
(2 kPa)
NOTE:
The Channel setup menu makes only those pressure labels available which
have been selected for the monitoring session in the IBP setup menu (or the default GP1/
GP2).
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...