5-22
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Temperature (T)
Message
Possible Cause
Suggested Action
Message: T >UL
<value>
Message: T <LL
<value>
The patient's temperature falls
outside the current temperature
alarm limits.
• Observe the patient.
• Change the alarm limits for this
patient.
• If you suspect a temperature
probe failure, check and replace
it.
Message: MultiMed
Disconnected
Parameter value:
<blank>
The M
ULTIMED
/N
EOMED
pod is
disconnected from the monitor.
• Verify the pod connections.
Message: T Fault
Parameter value: ***
The monitor has detected a
hardware failure.
• Turn the monitor off, then on.
• If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T
Unplugged
Parameter value: ***
The temperature probe is
disconnected from the M
ULTIMED
/
N
EOMED
pod.
The M
ULTIMED
/N
EOMED
cable is
defective.
The probe is defective.
The probe is not recommended for
use with your monitor.
• Verify the probe connection to the
pod.
• Verify the probe. If defective,
change the probe.
Message: T Test Fail
The monitor has detected a
problem in the temperature sensor
during startup.
• If the test fails, turn the monitor
off, then on.
• If the message does not clear,
take the unit out of operation and
call your Biomed.
Message: T Too Low
Message: T Too High
Parameter value:
<blank>
The detected temperature is
outside the measurement range.
• Check your patient.
• Check the probe for defects.
• Check the patient cable
connections.
• Change the probe, if necessary.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...