Non-Invasive Blood Pressure (NBP)
Message
Possible Cause
Suggested Action
Message: NBP Cannot
Measure
Parameter value: ***
Faulty equipment.
Improper cuff placement.
Patient’s pulse too low.
Excessive patient movement.
• Check the patient.
• Check the hose and cuff.
• Move the cuff to a limb with less
movement.
• Check proper placement of the
cuff.
• Restart measurement.
Message: NBP Artifact
Parameter value: ***
The patient is moving or shivering.
• Check the patient.
• Move the cuff to a different limb
with less movement.
Message: NBP No
Pulsation
Parameter value: ***
Weak signal. Monitor is unable to
detect a sufficient number of
pulsations of adequate amplitude
within two minutes.
• Check the patient and treat if
necessary.
• Check the hose and cuff.
• Check for proper size, placement
of cuff.
Message: NBP Cuff
Leak
Parameter value: ***
The monitor has detected a
significant drop in the cuff pressure
during inflation.
• Check for leaks in the cuff or
hose.
• Change the cuff and hose, if
necessary.
Message: NBP Fault -
Power Cycle Monitor
Parameter value: ***
Faulty equipment.
Improper cuff placement.
Transducer needs calibration.
NBP hardware needs repair or
calibration.
Hardware safety timer has expired.
• Check the hose and cuff.
• Check proper placement of cuff.
• Turn the monitor off, then on if
you detect a hardware failure.
• If the message does not clear,
take the unit out of operation and
call your Biomed or
DrägerService.
Message: NBP Mean
Only
Parameter Value:
NBP M <value>
NBP S <***>
NBP D <***>
The patient's pulse is too low for
the monitor to derive the systolic
and diastolic pressure values but
large enough to report the mean
pressure value.
Improper cuff placement
• Check the patient.
• Check proper placement of the
cuff.
Message: NBP Meas.
Time-out
Parameter value: ***
Measurement time has exceeded 2
minutes.
• Repeat the measurement.
Message: NBP Open
Line
Parameter value: ***
There is no significant increase in
cuff pressure during the inflation
cycle.
• Verify the cuff connection.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...