8-10
Infinity Vista
VF6
8 ECG
AND
H
EART
R
ATE
One- or Two-Channel Signal Processing
The monitor can process one or two ECG leads. If you select two-channel signal
processing, ECG and arrhythmia monitoring is less vulnerable to artifact, because the
monitor can assign a greater weight to the “cleaner” channel or even exclude a
channel from signal processing altogether, if it exceeds a certain level of artifact. In
this case, ECG monitoring continues without interruption, unless both channels start
showing excessive artifact.
The monitor processes the two leads displayed in the top two waveform channels. If
no lead or only one lead is currently displayed in these channels, the monitor uses the
last leads previously displayed there.
STEPS: Selecting One- or Two-Channel Signal Processing
1. Click on the
HR
parameter box.
2. Click on
More...
.
3. Click on
ECG Processing
.
4. Select the desired setting (
ECG1
or
ECG1&2
) and click the knob.
NOTE:
When you call up the second page of the HR Setup menu by clicking on
More...
(see below), the two ECG leads available for signal processing appear to the
right of the menu. To change these leads, exit the HR Setup menu, click on waveform
channel 1 or 2, and select a different ECG lead for display.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...