VF6
Infinity Vista
8-3
If using pre-gelled electrodes, verify that there is enough gel in the gel-filled area.
Preparing the Patient’s Skin
The quality of ECG monitoring depends largely on the strength and quality of the
signals received by the electrodes. Careful skin preparation and application techniques
assure strong signals with minimal artifact and interference.
Select flat, non-muscular sites to position the electrodes and follow the clinical
techniques of your hospital. We suggest the following standard technique:
1. Prepare the skin by clipping or shaving excess hair.
2. Remove any skin residue or oils with an alcohol pad.
3. Remove the outer epidermal layer as required to reduce skin impedance.
Mildly abrade only the electrode contact site with ultrafine sandpaper (220-
400 grit). Apply the electrodes one at a time and make sure the electrode gel is
in contact with the abraded skin area.
4. For severely diaphoretic patients, use a benzoin prep for tighter adherence of
the electrodes.
5. Inspect the electrode gel to make sure it is moist. Apply the pad with a circular
motion on the adhesive area first, then press gently on the gel area to prevent
the gel from being squeezed out.
6. Change electrodes every 24-48 hours to ensure a good quality signal.
However, if the electrocardiographic pattern becomes less distinct, if the
patient is diaphoretic, or if skin irritation develops, the electrodes must be
changed and reapplied sooner.
NOTE:
Never use disposable electrodes after their expiration dates or when gel has
dried out.
WA R N I N G : C o n d u c t i v e pa r ts o f e l e c t r o d e s a n d c o n n e c t o r s
( i n c l u d i n g t h e n e u t r a l e l e c t r o d e ) s h o u l d n o t c o n ta c t o t h e r
c o n d u c t i v e m a t e r i a l , i n c l u d i n g e a r t h .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...