10-2
Infinity Vista
VF6
10 ST S
EGMENT
A
NALYSIS
Overview
The monitor provides an option for ST segment analysis, available for the adult and
pediatric monitoring modes. When activated, this option analyzes the patient’s normal
QRS pattern and determines values for ST segment deviation (i.e., how far the ST
segment of the QRS complex is above or below the isoelectric line). ST segment
analysis is performed on the two leads selected for ECG monitoring. The default leads
for ST analysis are lead II and V (with a 5- or 6-lead cable), the same default leads
assigned for Arrhythmia monitoring. Upper and lower limits are user-selectable and
the monitor triggers an alarm in case of limit violation.
When you connect the monitor to a patient, the monitor begins the process of learning
the patient’s dominant ECG beat pattern: this pattern serves as a baseline for
determining subsequent normal beats. The monitor stores an average ST complex that
is derived from those beats classified as normal. In order to obtain an ST segment
deviation value, the monitor then compares the isoelectric and the ST segment level of
the average complex. The ST segment deviation represents the difference in voltage
between the isoelectric and the ST segment level of the average complex. The monitor
displays the ST segment deviation in the ST parameter box. The value is updated
every 15 seconds.
NOTE:
z
You can initiate a Relearn of the ECG reference pattern manually at any time
(see the chapter
Arrhythmia
).
z
ST deviation values are displayed in mm (default) or mV. Changing the units of
measurement is a password-protected function. For information, contact your
Biomedical personnel.
WA R N I N G : T h e S T a l g o r i t h m h a s b e e n t e s t e d f o r t h e a c c u r a c y
o f d e t e c t i n g S T s e g m e n t d e v i a t i o n s . T h e s i g n i f i c a n c e o f t h e
S T s e g m e n t d e v i a t i o n s m u s t b e d e t e r m i n e d b y a c l i n i c i a n .
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...