C
LEANING
AND
D
ISINFECTING
VF6
Infinity Vista
B-3
Reusable ECG Electrodes
z
Clean grabber-wire clips regularly with a toothbrush.
z
Remove any gel residue from the electrode by brushing it off under running
water.
z
Clean the electrodes with a gauze moistened with a soap solution.
z
Dry thoroughly with a lint-free cloth.
z
Disinfect the electrodes with a gauze moistened with diluted alcohol.
z
Dry thoroughly with a lint-free cloth.
Reusable SpO
2
Sensor
See the cleaning instructions and recommendations provided with the sensor.
NBP Cuff
Wipe the NBP cuff with a cloth moistened with soap and water or a solution based on
household bleach (1:10), alcohol, or phenol.
Temperature probes and cables
z
Do not use excessive pressure or flex the cables as this can stretch the
covering and break the internal wires
z
Clean the probes with a 3% hydrogen peroxide or 70% alcohol
z
Quickly immerse the cables in a detergent solution
z
Make sure the probe’s tip is firmly connected.
CAUTION:
The NBP cuff can be immersed in cleaning solution, but do not allow the
solution to enter the NBP hose. Warranty claims can be forfeited if cleaning solution is
allowed to enter the hose or the cuff.
CAUTION:
z
Do not use phenol disinfectants because vinyl absorbs them. Do not use strong
aromatic, chlorinated, ketone, ether or ester solvents. Do not immerse the
cables for any prolonged period in alcohol, mild organic solvents, or highly
alkaline solutions.
z
Never boil or autoclave the cable. Vinyl withstands temperatures up to 100 °C
but begins to soften at around 90 °C. Handle gently when hot and wipe away
from the tip toward the cable.
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...