O
VERVIEW
VF6
Infinity Vista
D-7
Monitoring Specifications
ECG
Accessories
3-,5-, and 6-Lead patient cable
Input Connector
2-pin connector to MultiMed pod
Input Parameters
3-Lead (RA, RL, LA and LL electrodes)
5-Lead (RA, RL, LA, LL and CHEST electrodes)
6-Lead (RA, RL, LA, LL and 2 CHEST electrodes)
Number of Channels
2
Sweep Speed
25 mm/s
Resolution
1/min.
Measurement Range
15 to 300 1/min.
Measurement Accuracy
±5 beats/minute or ±2%, whichever is greater
Response Time
< 7 seconds for a step change of 80 to 120 1/min.
< 13 seconds for a step change of 80 to 40 1/min.
• Response time is measured from the onset of the first QRS
complex of the new rate to the time the measurement reads
a value that is the original rate plus 63% of the change.
Report Interval (nominal)
2.1 seconds
Notch Filter Frequency
50/60 Hz
Monitor ±3 dB Bandwidth
0.5 Hz to 28 Hz (50 Hz)
0.5 Hz to 40 Hz (60 Hz)
Dynamic Range AC
±10 mV minimum
Dynamic Range DC
± 1 V minimum
Resolution
10.4
P
V
Electrode Off Impedance
> 100 M
:
Electrode On Impedance
< 40 M
:
Calibration Spike
1 mV ±5% amplitude, 20 msec duration
Defibrillation Protection
In accordance with IEC 601-2-27
Degree of Protection Against
Electric Shock
CF
Pacer Pulse Detection
On active ECG lead
Summary of Contents for Infinity Vista
Page 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Page 2: ......
Page 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Page 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Page 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Page 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Page 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Page 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Page 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Page 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Page 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Page 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...