4.1. Pump
•
Check oil level daily.
•
Please follow manual of pump manufacturer.
•
For oil change interval see manual. The oil change has to occur every 3 months.
4.2. Safety Components
•
Check following components semi-annual (see circuit diagram):
6.2.1. door switch
6.2.2. vacuum switch
6.2.3. relay on PCB
6.2.4. timer
6.2.5. valves
4.3. Cleaning
•
For the cleaning of the outside of the plasma system use only a light
wet cloth. For the cleaning of the chamber see chapter 1.4 Vacuum
Chamber. Never use solvents for the cleaning!
•
Stainless steel recipients are easier to clean with chrome polisher. Therefore remove /
uninstall electrode previously.
4.4. Seals
•
The door seal should be wiped off every day and semi-annual changed.
4.5. Electrode Retainer
•
Please test / check electrode retainer for leaks:
•
Pump off machine
•
Start plasma process
•
Wet electrode retainer with ethanol
•
The plasma colour should not change from purple into light blue.