10. The operating mode of machine is according to installed control (semi-automatic,
automatic or PC – control). Find more information to the controls in chapter 1.7 Controls.
As a matter of principle read following procedure for the plasma treatment:
a) Load chamber.
b) Press cover / door against chamber opening.
Important:
pay attention to the door switch. The door switch has to build a contact
with the door, because the generator can not ignite without this contact (safety
switch).
c) Turn on pump (’Pump’ button), press door against chamber opening until it will be
primed.
d) Adjust time (main interval at the back of the machine, process time at the
potentiometer at machine front). More information in chapter 1.7 Controls.
e) Turn on gas supply (’Gas’ button)
f) Adjust desired gas flow or pressure.
g) Adjust the desired generator power. At semi-automatic and automatic machines,
the power will be adjusted via a potentiometer. The upper row of numbers shows
the tenth, the lower the unit place. See figures below:
100%
power
65%
power
h) Start generator (The generator turns off automatic after cycle of time.)
i) Turn off pump, by pushing of ‚Pump’ button.
j) Vent chamber („Vent“ or „Flood“)
k) When chamber is ventilated, the door let’s easily open/remove.
The door switch is located next to the door seal.
Mostly on the
left side
horizontal.
Summary of Contents for femto