Ch. 6 - Calibration and Measurement
Measurement
User's Guide to the SPOT Cooled Color Digital Camera, 6/9/98
106
9.
Click on the Calibrate button. The number of pixels per 5 microns appears.
10.
Click on OK. Repeat steps 1 - 9 for each objective.
Measurement
Once you have specified the calibration for an objective, you can then use the SPOT
software to perform different types of measurements on an image.
To perform a measurement on an image, follow these steps:
1.
From the View menu, select the Show Measurements option (or press
[Ctrl M]
). The
Measurement window appears, as shown below:
2.
Select a Calibration Setup from the list. Typically, the Calibration Setups correspond
to your microscope’s objectives.
3.
Select from one of four measurement options:
−
Straight line
−
Curve
−
Angle
−
Region Area
!
Note: Angle measurements are independent of the Calibration Setup.
4.
Draw the line, curve, angle or area that you want to measure.
a.
Position the cursor on the image that you are measuring.
b.
Left click on the mouse. The drawing function is now enabled. Drag the mouse
across the image to perform the measurement.
c.
Left click at the point where you want to end the measurement.
d.
The length, number of degrees, or square units appears at the bottom of the
window.