6. REMOTE OPERATION
Page 151
© 2009 DH Instruments, a Fluke Company
HS
Purpose
Read or set the automated pressure control hold limit as a pressure value.
Command
Query
“HS
hold
”
“HS?”
Classic
“HS=
hold
”
“HS”
Arguments
Hold:
The hold limit in the current pressure unit of measure.
Remarks
The hold limit can be read and set as a pressure.
Example
(enhanced)
Cmd sent:
Query reply:
“HS .1”
“0.1 kPa”
(No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“HS? .1”
“0.1 kPa”
Example
(classic)
Sent:
Reply:
“HS=0.1”
“0.1 kPa”
Errors
ERR# 6
The
‘hold’
argument was invalid.
See Also
“HS%”
3.2.2 , 3.2.13
HS%
Purpose
Read or set the automated pressure control hold limit for as a percent of range span.
Command
Query
“HS%
hold
”
“HS%?”
Classic
“HS%=
hold
“
“HS%”
Arguments
hold:
The hold limit in % span of the active range.
Remarks
The hold limit can be read and set as a pressure or as a percent of the span.
Example
(enhanced)
Cmd sent:
Query reply:
“HS .01”
“0.0100 %”
(No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“HS? .01”
“0.0100 %”
Example
(classic)
Sent:
Reply:
“HS=.01”
“0.0100 %”
Errors
ERR# 6
The
‘hold’
argument was invalid.
See Also
“HS”
3.2.2 , 3.2.13
ID
Purpose
Read or set the user defined instrument identification label.
Command
Query
“ID string”
“ID”
Classic
“ID=string”
“ID”
Arguments
String:
An alphanumeric string up to 12 characters long.
Remarks
The user defined ID label can be used to allow the user to “tag” the PPC4 with a unique identifier.
This ID is stored in non-volatile memory and cannot be erased by a power failure, system fault or
reset
Example
(enhanced)
Cmd sent:
Query reply
“ID PPCK #A01”
“PPCK #A01”
(No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply
“ID PPCK? #A01”
“PPCK #A01”
Example
(classic)
Sent:
Reply:
“ID=PPCK #A01”
“PPCK #A01”
Errors
ERR# 6
The
‘string’
argument was longer than 12 char.
See Also
3.2.27.1
Summary of Contents for PPC4
Page 70: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 58 Notes...
Page 194: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 182 Notes...
Page 226: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 214 Notes...