[Limits]
from the
[Settings], <Control>, Pressure Control
screen or from the
control mode
shortcut open this screen as well.
Three fields are available to edit:
Hold Limit, Stability Limit
(see Section 3.2.13)
and
Jog Step
(see Section 3.2.20).
Edit the values of the desired parameters, noting the units of measure.
[OK]
to
accept and return to the Main Run Screen.
[Back]
to return to
<Control>
with no
changes.
[Esc]
or
to exit to the Main Run Screen with no changes.
To restore the default values of the three parameters use
[Default Limits]
.
Note
Custom control parameters are range specific. Changes made
in one range are stored for that range and are recalled when that
range is returned to. Changes made in one range do not affect
any other range.
5.7.8.3 Jog Step
PURPOSE
Adjust the nominal change in pressure caused by using the jog direct pressure
control keys (
and
) (see Section 3.2.20).
OPERATION
[Settings], <Control>, Jog Step
accesses the Jog Control Limits screen, with the
Jog Step field active.
Enter
a new value in the currently active pressure units.
[OK]
to accept the change and return to the Main Run Screen.
[Back]
to return to
<Control>
without making changes.
[Esc]
or
to exit to the Main Run Screen
without making changes.
5.7.9
Remote
PURPOSE
To configure the PPC4 COM1, COM2, IEEE-488 (GPIB), or LAN communication ports. To
test COM1 and COM2 communications. To select remote programming communications
format (see Section 3.2.24).
OPERATION
[Settings], <More >>>, <Remote>
opens the Remote settings screen.
The remote setting screen varies, depending on the communications hardware installed. The
following choices are always available:
COM1:
To view and edit COM1 settings (see Section 5.7.9.1).
COM2:
To view and edit COM2 settings (see Section 5.7.9.1).
Command Format:
To select remote programming command format (see Section
5.7.9.2)
If no optional interface hardware is installed, the following is also displayed.
USB:
Indicates front USB port only, not selectable
If the optional IEEE-488 hardware is installed, the following two options are displayed.
IEEE-488
:
To view and edit IEEE-488 settings (see Section 5.7.9.3)
USB:
Indicates front USB port only, not selectable