5. ADVANCED USER INTERFACE
Page 115
© 2009 DH Instruments, a Fluke Company
OPERATION
[Settings], <Pressure>, Resolution
opens the Resolution screen. The shortcut to
Resolution is the measured pressure field from the Main Run Screen (see Figure
11). In the resolution window choose from the available choices between 1.0% of
full scale and 0.0001% of full scale. The currently active resolution value is
shown in gray.
[Back]
to return to
<Pressure>
.
[Esc]
or
to exit to the Main Run Screen
without making a change.
Note
•
The resolution setting is range specific. A resolution
setting made in one range does NOT affect other ranges.
•
The measured pressure resolution is fixed for altitude units at
1 m in meters and 1 ft in feet.
•
The maximum resolution setting is limited in AutoRange
ranges that are less than 10% of the Q-RPT default full scale.
•
The maximum resolution setting when using a utility sensor
is 0.001% of utility sensor span.
5.7.7
Preferences
Use
[Settings], <Preferences>
to access
Screen Saver, Sounds, Time, Language
and
Security
. The
[Settings], <Preferences>
tab displays the current values of each of these
settings.
Select
and
Enter
to edit settings.
5.7.7.1 Screen Saver
PURPOSE
To adjust the PPC4’s Screen Saver settings (see Section 3.2.26.1).
OPERATION
[Settings], <Preferences>, Screen Saver
opens the Screen Saver screen. Edit
Time
, from 10 minutes to 8 hours. If Time is set to Off then the screen saver
function will be disabled and the screen will remain on as long as the PPC4 power
is on. Editing
Style
chooses the display when the screen saver is activated. The
choices include “Moving Pressure” which displays the current measured
pressure, moving on the screen, and “DHI” (the DHI logo).
[OK]
to accept and return to the Main Run Screen.
[Back]
to return to
<Preferences>
.
[Esc]
or
to return to the Main Run Screen without making a
change.
5.7.7.2 Sounds
PURPOSE
To adjust or suppress the PPC4 key press sound (see 3.2.26.2).
OPERATION
[Settings], <Preferences>, Sounds
opens the Sounds screen. From here choose
None, Low, Medium
or
High
for the valid key press frequency. The currently active
frequency is shown in gray.
Choose
None
to suppress the key press tone.