AUTOZERO
Purpose
Read or set the status of the AutoZ function.
Command
Query
“AUTOZERO
n
”
“AUTOZERO?”
Classic
“AUTOZERO=
n
”
“AUTOZERO”
Default
“AUTOZERO 1”
Arguments
n
:
’0’ Autozero OFF
’1’ Autozero ON
Remarks
The PPC4 AutoZ function can be turned ON and OFF. There is a separate AutoZ flag
for the gauge and absolute modes for each RPT. This command sets the AutoZ status
for the current mode of the active internal or external RPT only.
Example
(enhanced)
Cmd sent:
Query reply:
“AUTOZERO 0”
“0”
(No reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“AUTOZERO? 0”
“0”
Example
(classic)
Sent:
Query reply:
“AUTOZERO=1”
“AUTOZERO=0”
Errors
ERR# 6:
The argument was other than a ‘0’ or a ‘1’.
See Also
“ZOFFSET”, “MMODE”
3.2.22
CALAMB
Purpose
Read or set the on-board barometer calibration coefficients.
Command
Query
“CALAMB
adder, mult, CalDate
”
“CALAMB?”
Classic
“CALAMB
=adder, mult, CalDate
”
“CALAMB”
Defaults
“CALAMB
= 0.0, 1.0, 19800101”
Arguments
Adder:
The Barometer calibration adder (PA).In Pascal.
Mult:
The Barometer calibration multiplier (PM) from 0.1 to 100.
CalDate:
The date of the calibration in the format “YYYYMMDD”
Remarks
The barometer calibration information can be accessed with this program message. Using
this program message overwrites the current calibration coefficients, so caution must be
used. Changes made take effect immediately.
Example
(enhanced)
Cmd sent:
Query reply:
“CALAMB? 2.1, 1.000021, 20011201”
“ 2.10, 1.000021, 20011201” (No reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“CALAMB=2.1, 1.000021, 20011201”
“ 2.1, 1.000021, 20011201”
Errors
ERR# 6:
One of the arguments is out of range.
See Also
3.2.27.3, 7.3
Summary of Contents for PPC4
Page 70: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 58 Notes...
Page 194: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 182 Notes...
Page 226: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 214 Notes...