PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 8
2.3.2.2 Rear Panel
ATM port
COM2 Connector
Optional USB Connector
Optional Ethernet Connector
Optional IEEE-488 Connector
COM1 Connector
Power Switch
Fuse
Electrical Power Connector (IEC-320-C13)
Product Label
Drivers (12 V) Connector
Pressure Connection, TEST(+)
Pressure Connection, TEST(-)
Pressure Connection, SUPPLY
Pressure Connection, EXHAUST
Figure 3.
Rear panel with optional connectors
2.3.3
Power Connection
Check that the PPC4 power switch is OFF.
Connect the supplied power cable to the rear panel power module.
Connect the other end of the power cable to an electrical supply of 100 to 240 VAC,
50-60 Hz.
2.3.4
Connecting To A Pressure Supply (Supply Port)
Using a pressure connecting hose or tube of appropriate pressure rating, connect the
pressure supply to the
SUPPLY
port on the rear panel of PPC4. The PPC4
SUPPLY
port
connection is
1/8 in. NPT female
.
The supply pressure should be equal to the maximum PPC4 control pr 10 % (or
70 kPa (10 psi) for a BG15K Q-RPT). Lower gas pressure sources can be used but should
exceed the maximum desired test output pressure by 10 to 20 %.