Delay at Target:
Enter
the time at Max Target and Min Target before
continuing with the next step in the sequence. If Control
at Target is set to
Yes
then pressure control will be
maintained during this delay. If Control at Target is set to
No then pressure control will abort and the PPC4 will be
idle during this delay.
[Back]
returns to the AutoTest screen.
[Esc]
or
exits the Exercise menu and
returns to the Main Run Screen.
[OK]
accepts the settings and proceeds to the
Exercise Run Screen.
The Exercise Run Screen displays the Cycle count (cycle in progress / total
number of cycles) and the Time at Ready which counts down from the Delay at
Target value to 0.
In the Exercise Run Screen,
[Start]
to begin the Exercise sequence. At any time,
[Stop]
to abort the sequence and reset the counter.
[Back]
aborts the sequence
and returns to the Exercise entry screen.
[Esc]
or
aborts the sequence and
returns to the Main Run Screen.
When the exercise sequence is complete the PPC4 sounds three two second
beeps and returns to the main menu.
5.7.2.2 Quick Test
OPERATION
To run a Quick Test sequence,
[Tools], <AutoTest>, Quick Test
from the Main
Run Screen.
From the Quick Test screen enter the following:
Max:
The maximum test pressure for the sequence. The
value must be within the current AutoRange limits (see
Section 3.2.8). To change the current AutoRange see
Section 5.7.1.
Note
The maximum pressure of the currently available AutoRange is
shown in the upper left hand corner of the Quick Test window.
See Section 5.7.1.
Min:
The minimum test pressure for the sequence.
Step:
The step size in % of test sequence span (Max-Min).
Test will run with pressure steps of (Max-Min)*Step%
with the uneven increment at the Max pressure if
applicable.
Direction:
Direction of test execution.
Up
– starts at Min and ends at Max
Down
– starts at Max and ends at Min
Up/Down
– sequence runs from Min to Max and back
to Min
Down/Up
– sequence runs from Max to Min and back
to Max