PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 98
[OK]
from the AutoRange Confirmation screen creates the range,
[Back]
returns to the AutoRanger
screen and
[Esc]
or
exits to the Main Run Screen and does not create the AutoRange.
[Save Range]:
Click to save an AutoRange to the
Range
drop down list.
[Delete Range]:
Click to delete the currently selected AutoRange from the
Range
drop down list.
[RPT Search]:
To search for, initialize and display internal and external Q-RPTs available to the
PPC4 system. Selecting an available Q-RPT changes Range to the default range of that Q-RPT.
Note
•
[Esc]
from the RPT Search screen the PPC4 will return to the Main Run
Screen and no AutoRange changes that were made after the last
[Save]
will be stored.
•
[OK]
creates the newly defined AutoRange and to uses it in operation.
If a range is selected from the list,
[OK]
changes the active range to the
selected Range.
[Esc]
returns to the Main Run Screen without
changing the currently active AutoRange.
•
To verify or confirm the range from the Main Run Screen, use
[AutoRange]
from the Main Run Screen and view the
Range
field to
view the active range (see Section 3.2.8) or view the AutoRange and
active Q-RPT spans and indicators from the AutoRanger bar on the
main run screen (see Figure 14).
The AutoRange Bar in the middle of the PPC4 Main Run Screen displays the range and a
graphical indicator of the measured pressure within the range for both the active AutoRanged
Range and the active Q-RPT (see Figure 14, showing a 200 kPa AutoRanged Span on a
700 kPa Q-RPT).
AutoRange Bar
AutoRanged Span Indicator
AutoRanged Span
Q-RPT Span
Q-RPT Span Indicator
Figure 14.
AutoRange Bar
Caution
•
Before operating PPC4 with external Q-RPTs, check carefully that the
Q-RPT TEST(+) port is connected to the PPC4 TEST(+) port. If the
external Q-RPT is NOT connected to the PPC4 TEST(+) port, it will not
measure the pressure generated by PPC4 and PPC4 may reach a
pressure higher than the target pressure before its watchdog function
recognizes that the external RPT is not connected.