PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 64
MENU SELECTION
SEE
SECTION
<Spec>, <Cal>, <Lo RPT>, <View/Edit>
7.2
<Spec>, <Cal>, <Barometer>, <View/Edit>
7.3
<Spec>, <Log>
4.5.9
<ARange>, <Abs/Gage/Neg Gage>
4.5.10
<Leak>, <Run/View>
4.5.11
<RPT>
4.5.12
<Unit>
4.5.13
<Range>
4.5.14
<Mode>, <Abs/Gage/Neg Gage>
4.5.15
<Head>
4.5.16
<Purge>
4.5.17
4.4
Menu Tree Operation
PURPOSE
Note
from the Main Run Screen to access all menu selections in the Basic Interface.
<Setup>
accesses a menu of functions and features commonly used in setting up and using PPC4.
4.4.1
<Setup>, <Range>
PURPOSE
Save AutoRange ranges and associated settings for recall. Delete previously saved
AutoRange ranges.
4.4.1.1 Saving An AutoRanged Range
OPERATION
To save a range created by AutoRange, the range must be the active range.
Use AutoRange to create the range (see Sections 3.2.9 and 4.5.10) and then
make any desired feature and setting adjustments.
Once the AutoRanged range is set up as desired, use
<Setup>
,
<Range>
,
<Save>
.
The display is:
1.
Indication that this is an AutoRange range to
be saved.
2.
Designator and position of the Q-RPT used by
the range to be saved.
3.
Pressure unit of measure of the range to be
saved.
4.
Full scale pressure of the range to be saved in
gauge mode and absolute mode if available.
Save range A700K X1H
kPa + 300g/100a