4. BASIC USER INTERFACE
Page 65
© 2009 DH Instruments, a Fluke Company
to save the range or
to return to the Main Run Screen without saving the
range.
4.4.1.2 Deleting AutoRange Ranges
OPERATION
To delete a single saved AutoRange range, use
<Setup>
,
<Range>
,
<Delete>
.
If there are any saved ranges available to delete, the display is:
1.
Indication that the cursor control keys may be
used to toggle through the saved ranges to
select the one to be deleted.
2.
Designator and position of the Q-RPT used by
the range to be deleted.
3.
Pressure unit of measure of the range to be
deleted.
4.
Full scale pressure of the range to be deleted
in gauge mode and absolute mode if available.
< and > A700K X1H
kPa + 300g/100a
Use the cursor control keys to toggle through the ranges available to be deleted.
When the desired range to be deleted is displayed,
. If you are sure you want to
delete this range,
again in response to the
<Delete?>
query.
to return
to the AutoRange
screen without deleting a range.
To delete all of the saved AutoRange ranges, press
<Setup>
,
<Range>
,
<Delete
All>
. If you are sure you want to delete all the ranges, respond
<Yes>
to the
<Delete all saved ranges?>
query. If not, respond
<No>
.
4.4.2
<Setup>, <Res> (Resolution)
PURPOSE
To set the resolution with which measured pressures and other pressure indications and
settings are displayed (see Section 3.2.19).
OPERATION
To access the resolution function, select
<Setup>
,
<Res>.
The
display is:
Display resltn: IH
0.0010 %FS < and >
Use the cursor control keys to set the desired level of resolution.
to return to the Main
Run Screen with the new resolution setting active or
to make no changes.
4.4.3
<Setup>, <Jog>
PURPOSE
Adjust the nominal change in pressure caused by using the jog direct pressure control keys
(
and
) (see Section 3.2.20).
OPERATION
To access the jog step adjustment function, select
<Setup>
,
<Jog>
. The display is:
Jog step size: IH
0.020 kPa
Edit the value of the jog step as desired.
to return to the Main Run Screen with the new
jog step size active.
to return to the Main Run Screen with no change to the jog step size.