PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 168
SS%
Purpose
Read or set the current stability limit as a % of range.
Command
Query
“SS%
n
”
“SS%?”
Classic
“SS%=
n
”
“SS%”
Arguments
n:
The stability limit in %FS of the current active range.
Remarks
The stability limit can be read and set as a percent of the full scale range of the Q-RPT range. If this
program message is used to set the stability limit, the PPC4 will then use CUSTOM control settings.
Example
(enhanced)
Cmd sent:
Query reply:
“SS% .1”
“0.10 %” (No reply from IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“SS%? .1”
“0.10 %”
Example
(classic)
Sent:
Reply:
“SS%=.1”
“0.10 %”
Errors
ERR# 6
The argument was invalid.
See Also
“SS”
3.2.13
SS
Purpose
Read or set the current pressure stability limit.
Command
Query
“SS
n
”
“SS?”
Classic
“SS=
n
”
“SS”
Arguments
N:
The stability limit in the current pressure unit of measure.
Remarks
The stability limit can be read and set as a pressure. The stability limit is used as the
Ready/Not Ready
criterion in static control mode and when PPC4 not controlling.
Example
(enhanced)
Cmd sent:
Query reply:
“SS .1”
“0.10 kPa/s”
(No reply from IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“SS? .1”
“0.10 kPa/s”
Example
(classic)
Sent:
Reply:
“SS=.1”
“0.10 kPa/s”
Errors
ERR# 6
The argument was invalid.
See Also
“HS”
3.2.13