PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 210
SYMPTOM
PROBABLE CAUSE
SOLUTION
Will not vent.
System is vented but does not indicate zero
because measurement mode is absolute.
Check measurement mode setting and
current value of atmospheric pressure if
absolute. 3.2.11
VENT
port is plugged.
Open
VENT
port to atmosphere.
Vent valve not operating.
Contact
DHI
Authorized Service Provider.
Table 36
Will not vent while purging. Stuck at
pressure during purging.
Purge function is not activated, Purge
function is activated but no SPLT is
connected or SPLT valve is not operating.
Activate SPLT; correct SPLT connection or
repair SPLT exhaust valve. 3.2.27.4,
2.3.7.1