PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 164
READYCK
Purpose
Read or set the
Ready
check flag.
Command
Query
“READYCK 1”
“READYCK?”
Classic
“READYCK=1”
“READYCK”
Remarks
The internal ready check flag is cleared whenever the PPC4 reaches a
Not Ready
(NR)
condition. The “READYCK” query returns the status of the flag. The flag is set by sending
the “READYCK 1” program message while the PPC4 is in a
Ready
condition. The
“READYCK” program message query can then be used at a later time to determine whether
a
Not Ready
condition has occurred since the ready check flag was set.
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK 1”
“1” (no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK?”
“1”
Example
(enhanced)
Cmd sent:
Query reply:
“READYCK?”
“1” (if PPC4 condition has stayed
Ready
)
“0” (if PPC4 condition has NOT stayed
Ready
)
Example
(classic)
Sent:
Query reply:
“READYCK=1”
“READYCK=1”
Example
(classic)
Sent:
Query reply:
“READYCK”
“READYCK=1” (if PPC4 condition has stayed
Ready
)
“READYCK=0” (if PPC4 condition has NOT stayed
Ready
)
Errors
ERR# 6:
Argument is not a ‘0’ or a ‘1’.
See Also
“SR”
3.2.3
REMOTE
Purpose
Lock out the front panel keypads during remote operation.
Command
“REMOTE”
Classic
“REMOTE”
Remarks
The PPC4 goes into remote mode whenever communications take place. The user can
return to local operation by pressing
. The REMOTE program message locks out the front
panel completely. The only way to unlock the front panel after the “REMOTE” command is
using the “LOCAL” program message, the IEEE-488 “GTL” command, or by cycling the
PPC4 power.
Example
(enhanced)
Cmd sent:
Reply:
“REMOTE”
“REMOTE” (no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“REMOTE?”
“REMOTE”
Example
(classic)
Sent:
Reply:
“REMOTE”
“REMOTE”