6. REMOTE OPERATION
Page 133
© 2009 DH Instruments, a Fluke Company
The next dialog defines the RS232
port to use for communication.
Select the host PC’s COM port
that will be connected to the
PPC4. Do not confuse this
selection with the COM1 and
COM2 ports on the back of the
PPC4.
After the COM port selection, the
RS232 settings of the PPC4 must
be defined. The specific PPC4
settings are accessed through the
PPC4 front panel. RS232
handshaking is not supported by
the PPC4, therefore None is the
appropriate selection for Flow
Control.
The focus will return to the main
Hyperterminal program after the
COM port properties are defined.
At this point, the last steps are to
define command and response
terminators. Access the [File]
menu and select [Properties]