PPC4™ OPERATION AND MAINTENANCE MANUAL
© 2009 DH Instruments, a Fluke Company
Page 110
5.7.5.3 Reset
PURPOSE
To reset various PPC4 settings to default or factory values. Use
[Tools],
<System>, Resets
to access the Reset functions (see Section 3.2.25).
Caution
PPC4 reset functions will change current settings to factory
defaults. These may include settings vital to PPC4 operation
and affecting the calibration of the quartz reference pressure
transducers (Q-RPTs). Reset functions should only be used by
qualified personnel with knowledge of reset consequences. Reset
functions should not be used experimentally.
5.7.5.3.1
Settings
PURPOSE
Sets most general operating parameters back to default values. Does not affect
calibration coefficients, remote interfaces or AutoRange ranges (see Section 3.2.25.1).
OPERATION
From the Main Run Screen use
[Tools], <System>, Reset, Settings.
[Yes]
to activate changes (perform the reset). The PPC4 will cycle power to
complete and return to the Main Run Screen.
[No]
to make no changes and
return to the Reset screen.
5.7.5.3.2
Units
PURPOSE
Resets Unit selections to factory defaults (see Section 3.2.25.2).
OPERATION
From the Main Run Screen use
[Tools], <System>, Reset, Units.
[Yes]
to activate changes (perform the reset. The PPC4 will cycle power to
complete and return to the Main Run Screen.
[No]
to make no changes and
return to the Reset screen.
5.7.5.3.3
AutoTest
PURPOSE
Resets Exercise, Quick Test and Advanced Test to default characteristics (see
Section 3.2.25.3).
OPERATION
From the Main Run Screen use
[Tools], <System>, Reset, AutoTest.
[Yes]
to activate changes (perform the reset. The PPC4 will cycle power to
complete and return to the Main Run Screen.
[No]
to make no changes and
return to the Reset screen.
Summary of Contents for PPC4
Page 70: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 58 Notes...
Page 194: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 182 Notes...
Page 226: ...PPC4 OPERATION AND MAINTENANCE MANUAL 2009 DH Instruments a Fluke Company Page 214 Notes...