5. ADVANCED USER INTERFACE
Page 101
© 2009 DH Instruments, a Fluke Company
[OK]
to accept test settings and continue.
[Back]
to return to
<AutoTest>
.
[Esc]
or
to exit to the Main Run Screen.
In the next screen enter the following:
Continue Mode:
From the drop down list choose User or Timer. If User is
chosen the test sequence will proceed to the next target
value only after
or the knob is pressed. If the Timer
is selected then the next test sequence will proceed after
the timer value has elapsed.
Timer:
The time, in seconds, the pressure will hold at each step
if the Timer mode is selected. Timer can be between 1
and 9999 seconds.
[OK]
to accept settings and continue to the Quick Test run screen,
[Back]
to
return to the Quick Test setup screen or
[Esc]
or
to exit to the Main Run
Screen.
In the Quick Test Run Screen,
[Start]
begins the Quick Test sequence. Once the
sequence begins, the Point counter displays current sequence increment / total
increments. If the control mode is dynamic (see Section 3.2.2.1), once a
Ready
condition has been achieved (see Section 3.2.3)
and
can be used to
adjust the pressure set by the PPC4 until the UUT reaches the cardinal value of
the pressure increment using the Jog function (see Section 3.2.20). In User
Continue Mode, once the UUT has reached the cardinal value of the pressure
increment then use
[Next Point]
. In Timer Continue Mode, once the pressure is
Ready
, the sequence will proceed automatically after the delay has elapsed.
Note
While the Quick Test is running, the upper half of the display is
exactly the same as the Main Run Screen, displaying the Ready
indicator, control indicator, measured pressure, target pressure
and pressure rate of change. See Section 5.1.
At any time during the sequence,
[Previous Point]
causes the PPC4 to go back
and retake the previous test point.
[Next Point]
stops the current point and
moves to the next test point in the sequence. On the last test point
[Stop]
is
shown instead of
[Next Point].
Completing the last point or
[Stop]
moves to the
Quick Test summary screen.
[Back]
aborts the sequence and returns to the previous screen.
[Esc]
or
aborts the sequence and returns to the Main Run Screen.
Once the test is complete the results are shown on the Quick Test summary
screen. The pressure values from the PPC4 are shown with the respective
sequence numbers. If there are more than eight points in the test then a scroll
bar is shown.
[Back]
returns to the Quick Test Run Screen to start the same test sequence
again.
[Esc]
or
returns to the Main Run Screen.