PARAMETER REVIEW
SUSPECT/
DEFINITIVE
FLAG WINDOW
TRANSMIT
MESSAGES
APPEAR HERE
POPULATION
CONDITION
MESSAGES
7188097A
FLOW CELL
CLOG
MESSAGES
Population Condition Messages
WBC, RBC, Plt, Retic
Condition messages describe the WBC (including Differential, if performed), RBC and Plt
populations. These messages occur in the lower left corner of the Run Samples screen. They indicate
the WBC, RBC, Plt and Retic population status only. If there is a voteout for WBC, RBC or Plt, no
message appears for the respective parameter. If CBC only is performed, population messages refer to
WBC, RBC and Plt counts only.
WBC Differential
A nonnumeric Diff percent result gives an
Abnormal WBC Population
message. The abnormal WBC
message may not signal only morphological abnormality. When paired with a non-numeric result
such as ....., or :::::, it is due to partial clogs, full clogs, incomplete computations. Verify results
according to your laboratory procedures. See Table 7.
PN 4237188A (December 1993)
83
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......