
SAMPLE ANALYSIS
4.
Cycle the sample:
a.
Immerse the aspirator tip in the sample
tube.
b.
Press and release the activator.
7188057A
IMPORTANT
Do not remove the tube from the aspirator tip until the STATUS message on the Analyzer CRT
changes from
INTRO SAMPLE to WIPE, and there is a beep.
WARNING
Be sure to remove the sample tube and be clear of the area around the aspirator tip before the
STATUS field displays the
PRIME message.
c.
Remove the tube of primer when
•
you hear the beep and
•
the Analyzer CRT STATUS field
displays
WIPE
.
d.
Wipe the tip with a lint-free tissue.
5.
Verify sample ID numbers before you report
results.
66
PN 4237188A (December 1993)
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......