DAILY QUALITY CONTROL
10. If a control is outside limits, the system
•
stops if AutoStop is on (Status line
shows QC
↑
)
•
turns the QC
↑
symbol red
•
displays an error message
•
flags results with H (High) or L (Low).
DMS
clears the error messages before you
continue.
Check Table 3 for possible causes of errors.
CBC/DIFF CONTROL RUN
FILE:
4
LOT: 884300
OPR: CDS
SHIFT : 1
LEVEL: Normal
EXP DATE: 10/08/93
RUN:
5
10/01/93
07:25
RESULTS
ASSAY
DIFF
LIMIT
RESULTS
ASSAY
DIFF
LIMIT
WBC
9.9
9.3
0.6
0.7
RBC
5.46
5.44
0.02
0.18
HGB
16.1
16.2
-0.1
0.6
NE%
54.3
54.5
-0.2
5.0
HCT
47.9
47.3
0.6
2.7
NE#
5.3
5.1
0.2
2.0
MCV
87.7
86.9
0.8
3.0
LY%
30.1
29.3
0.8
5.0
MCH
29.4
29.8
-0.4
1.2
LY#
3.0
2.7
0.3
1.6
MCHC
33.5
34.2
-0.7
1.7
RDW
13.7
13.3
0.4
1.5
MO%
7.9
8.6
-0.7
3.0
MO#
0.8
0.8
0.0
0.8
PLT
236
239
-3
25
MPV
10.6
10.2
0.4
2.0
EO%
7.6
7.5
0.1
2.0
EO#
0.8
0.7
0.1
0.7
BA%
0.1
0.1
0.0
0.5
BA#
0.0
0.0
0.0
0.1
F1-Help
F2-File
F4-Print
F6-Graph
F8-Delete
F9-Menu
F12-Graphics
10/10/93 07:35
OPR
DMS
↑
PR
↑
TK
↓
HC
↑
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
22
4C
®
PLUS Cell Control and Commercial CBC-Only Cell Controls
CAUTION
Do not cycle any commercial CBC cell control or calibrator material in the CBC+DIFF or RETIC mode.
If you do, it will damage the flow cell.
1.
ANALYZER
MAIN
SYSTEM CONFIGURATION
MODE OF OPERATION
Press as needed to set to CBC.
MAIN MENU
PN 4237188A (December 1993)
41
Summary of Contents for STKS
Page 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Page 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Page 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Page 124: ...DMS BASICS 114 PN 4237188A December 1993...
Page 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Page 168: ...DATA BASE 158 PN 4237188A December 1993...
Page 204: ...APPENDIX A 194 PN 4237188A December 1993...
Page 216: ...APPENDIX C 206 PN 4237188A December 1993...
Page 228: ...APPENDIX D 218 PN 4237188B March 1995...
Page 230: ...REFERENCES 220 PN 4237188B March 1995...
Page 236: ......