SAMPLE ANALYSIS
1.
Check for
ANALYZER
•
STATUS: READY
•
ALERT: (is blank)
•
CBC
↑
or
CBC
↑
and DIFF
↑
If you need to change mode:
MAIN
SYSTEM CONFIGURATION
MODE OF OPERATION
Press the key as needed to set
test mode.
*SYSTEM RUN*
STATUS:
ALERT:
READY
NUMBER OF ASPIRATIONS/TUBE 01
(MESSAGE AREA)
MAIN
03/14/93
10:45:37
BD
CBC
DIFF
RETICS
7188050A
MAIN MENU
SYSTEM RUN
2.
Prime the instrument:
a.
DILUTER
PRIME
APERT
b.
Select
DMS
Sample Analysis
Run Samples
Other
.
Prime
to turn Prime on.
.
62
PN 4237188A (December 1993)