STARTUP
6.
Mix controls gently according to directions
in the package insert. Perform the 8/8/8
procedure twice.
8 TURNS
7022034B
8 TURNS
8 INVERSIONS
7.
Cycle the control:
a.
Immerse the aspirator tip in the control
vial.
b.
Press and release the activator.
c.
Remove the control when
•
you hear the beep and
•
the Analyzer CRT status field
displays
WIPE
.
d.
Wipe the tip with a lint-free tissue.
7188057A
8.
View the results on the CBC/DIFF
CONTROL RUN screen. To see WBC, RBC
and Hgb graphs of the last 10 cycles, press
.
DMS
Press
again to view the other parameter
graphs.
9.
To review control files, control graphics and
graphs, see the Controls chapter.
40
PN 4237188A (December 1993)