STARTUP
5.
When cycling is complete,
Other
.
DMS
to turn Prime off.
.
IMPORTANT
You must properly warm and mix 5C cell control. Follow the procedure on the package insert. To
maintain a high read rate, keep the bar-code label free of blood, scratches and powder from gloves.
6.
Prepare 5C cell controls according to
directions in the package insert.
7.
Put the control tubes
•
Normal
•
Abnormal I
•
Abnormal II
in a standard 5 mL cassette with the bar-
code labels facing up.
You can put controls and patient samples in
the same cassette.
7188041A
36
PN 4237188A (December 1993)