STARTUP
14.
Latex
DMS
15.
DILUTER
ENTER
to reactivate function 57.
POWER
ON
D R A IN
R IN S E
CLEAR
APERT
START/
CONT
POWER
OFF
START
UP
SHUT
DOWN
ALARM
RESET
P R E M IX
S T O P
7
8
9
4
5
6
1
2
3
C E
0
F
ID
EN TE R
PRIME
APERT
7 1 8 8 0 0 8 A
PRESS MANUAL
16. Gently mix the control. Invert the vial five to
eight times. Avoid foaming.
7188010A
17. a.
Immerse the aspirator tip in the
LATRON control vial.
b.
Press and release the activator.
c.
Remove the control when
•
you hear the beep and
•
the Analyzer CRT STATUS field
displays
WIPE
.
d.
Wipe the tip with a lint-free tissue.
7188017A
24
PN 4237188A (December 1993)